Colonoscopy results are not enhanced by use of magnet endoguide in specialist practice

Anders Bak-Christensen1, Elisabeth Knudsen2, Jakob Hendel3, Inge Bøtker-Rasmussen Ifaoui1, Lars Lehrskov-Schmidt4 & Lene Hendel2

1. jun. 2013
12 min.

Faktaboks

Fakta

Since the introduction of a magnetic positioning device for endoscopy [1], its significance for colonoscopy has been widely discussed, and several articles concerning its effect on learning curves, success rates, safety and patient comfort have been published [2-4].

However, the impact on colonoscopies carried out in a specialist practice with a highly experienced staff and a large amount of open access patients has not previously been reported [5].

MATERIAL AND METHODS

Patients

All patients referred to the two clinical units for colonoscopy were invited to participate in the study, acknowledging the exclusion criteria. Theoretical power calculations were non-applicable in this study, as it was designed to run for a two-year period during which we expected to include 1,000 patients. If no effect could be established within this time or volume frame, the equipment would be deemed of no importance in a basic clinical context.

In 16 months, 1,068 patients were referred for colonoscopy and 1,004 were found eligible and entered the study.

Reasons for exclusion are shown in Figure 1.

Randomisation

All patients were given verbal and written information and signed an informed consent document.

The patients were randomised for investigation with our standard equipment, with or without the endoguide system (unit of magnetic positioning device, UPD) connected. The investigational procedure – including the bowel cleansing procedure (see below) – was not otherwise altered. The randomisation was carried out as allocation by date.

The UPD alternation between the two clinics was decided and handled by the company administering and transporting the UPD. UPD alternation on the one hand and scheduling of patient appointments by the clinics’ secretaries on the other were carried out independently.

Equipment

The standard endoscopy equipment was supplemented with a portable magnetic endoscope imaging unit (Olympus ScopeGuide system, UPD) [6].

Colonoscopy

Bowel preparation: PICOPREP was routinely used for bowel cleansing. The over-the-counter (OTC) medication was distributed to the patient by the clinic free of charge, and the patient received written and verbal instructions on how to use it properly.

Sedation was optional and patient-decided.

After the procedure, the patient was invited to indicate discomfort during the procedure as a visual analogue scale (VAS) (VAS 1) score 0-10, 0 being no discomfort at all, 10 being severe discomfort (painful). The patient received a second, identical VAS form (VAS 2) to fill in at home the day after the colonoscopy and sent in a prepaid envelope to the clinic. The patient was carefully instructed (before the procedure) to record the discomfort of the investigation performed the previous day.

In a subgroup of patients, the endoscopy assistant also filled in a VAS to give a staff impression of how the patient reacted to the investigation (VAS 3).

The concept that the endoscopic procedure would be the same whether the patient participated in the study or not was thoroughly described to the patient, and written informed consent for VAS score estimation was obtained.

Statistical analyses

First, VAS scores and intubation time were compared between the UPD and the non-UPD group. As the variables were not normally distributed, we used the non-parametric Wilcoxon rank sum test and the Wilcoxon signed rank test for paired observations. The proportion of successful endoscopy for each physician as well as the total proportion were calculated and compared between the two groups by either the χ2-test or Fisher’s exact test, whichever appropriate.

Furthermore, we performed multivariate regression analyses on three different outcome variables. An ordinary logistic regression analysis was carried out on the binary outcome measuring success or failure. As the assumptions of a linear regression model were not fulfilled for the two continuous outcome measures, VAS and intubation time, we grouped the observations according to the percentiles of their distribution and performed ordinal logistic regression analyses. Gender, age and UPD were included as explanatory variables in all regression analyses. As age was not a linear predictor in two of the three regression analyses, we included this variable as categorical in all three regressions defining age groups according to the age distribution. We estimated robust standard errors in order to account for clustering of patients.

All analyses were performed using the statistical software Stata version 11 (Stata Corp LP, College Station, TX) and the level of significance was set at 5%.

Trial registration: The study was approved by the Danish Data Protection Agency (journal no. 2009-41-3716), by the National Ethics Committee (journal no.: H-1-2009-80) and registered with ClinicalTrials.gov (protocol no: NCT01055782).

RESULTS

A total of 1,004 patients out of 1,068 were found eligible to enter the study according to the protocol (Figure 1). Distribution of patients between the clinics, patient demography, distribution of use versus non-use of UPD, and success rates for the clinics as well as for the individual endoscopists are shown in Table 1 and Table 2.

In all, 126 patients were investigated without sedation. Also in this subgroup, the use of UPD had no significant influence.

A total of 42 of 1,004 colonoscopies were withdrawn for reasons that were not endoscopist-related (Figure 1).

Success rate

A total of 555 patients were investigated with UPD; 19 patients had to be withdrawn prior to or during the endoscopy. In all, 41 colonoscopies were not completed (failures) and 426 patients were investigated without UPD. Furthermore, 23 patients had to be withdrawn ​ prior to or during the endoscopy, and 41 colonoscopies were not completed (failures). Patient demography and diagnostic pattern in failures did not differ from the diagnostic referral pattern.

The UPD success rate was 92.4%, whereas the non-UDP success rate was 90.4%. This difference is statistically insignificant as were the success rates calculated for the individual endoscopists (Table 2). The proportion of ileal intubation is also presented in Table 1.

Visual analogue scale score

A total of 936 (93%) patients completed a VAS 1 form, 884 (88%) patients completed a VAS 2, and 432 (43%) patients completed a VAS 3 form. No significant differences allocated to the use of UPD were demonstrated. In both groups, VAS 2, however, was significantly lower than the corresponding VAS 1 (Table 3, Table 4 and Table 5).

Duration of procedure: The duration of intubation time (minutes) was compared with and without UPD and no significant differences were demonstrated (Table 3).

In Table 6 multivariate analyses for success rate, VAS-score and intubation time show no significant difference between UPD and non-UPD.

Adverse events: One patient was hospitalised (self-admittance) due to abdominal pain and discharged without intervention after spontaneous relief.

DISCUSSION

Quality assessment, quality control and accreditation are currently important issues.

In this context, access to specialised equipment is very much in play. Success rates and patient safety are punch words when planning colon cancer screening programmes. Hence, it was suggested that access to a positioning device during colonoscopy may increase patient safety and procedure efficacy.

However, the use of a UPD has previously been shown to favour less experienced colonoscopists and to be of no importance for highly experienced colonoscopists [7-9].

Our study confirms that access to a UPD does not significantly enhance the performance of experienced endoscopists in specialist clinics.

Although the statistics show no evidence of beneficial effects of the UPD, either in success rate or in patient comfort (VAS score), all investigators but one preferred working with it.

In the present study, we confirmed that a specialised stand-alone unit not located within a hospital is sufficiently equipped and has an adequate intake of colonoscopies per endoscopist to participate in colon cancer screening.

One of the participating clinics was staffed with surgeons, the other with gastroenterologists (internists). This is to some extent reflected in the referral diagnoses and, consequently, in the endoscopic diagnoses. The attitude towards ileocaecal valve intubation also differs between surgeons and gastroenterologists, but the reluctance of the former towards ileum intubation changed to acceptance during the study, which obviously brought on an increasing success rate of ileocolonoscopy, but as attempted ileal intubation was not mandatory, ileal intubation cannot be included in the statistical evaluation.

We found no individual correlation between the number of colonoscopies, success rate and VAS score in our study, but as a rule, all of the participating endoscopists completed more than 200 colonoscopies annually. This is in accordance with other publications [10-12].

However, it seems to be of great importance to constantly monitor one’s success rate. The single investigator (no. 5) who had a significantly lower success rate than the other investigators had previously demonstrated a much higher success rate, but obviously lost momentum during the study, while all other investigators were performing consistently.

Previously reported success rates vary widely [7-9, 11-13].

It is not simple to compare results. The methods used for recording of success rate and for determination of patient discomfort differ considerably between studies. Some papers record calls for assistance during a colonoscopy, other papers deal with a “crude” patient population, as we do, and others with a highly cleansed patient material.

Dechêne et al [9] looked into different components of patient discomfort with emphasis on turns of the patient and abdominal compression and found the latter to be significant for patient comfort and dependent on the use of a magnet endoguide. In our patient material, we recorded only intubation time and patient discomfort and found no difference between the two patient groups.

The variety of diagnoses in patient populations from various studies and from hospital and practice settings may differ, although probably not as much as may be expected. There are no differences in diagnoses when we compare failed and successful endoscopies.

Our scope was primarily to compare the efficacy of having access to a UPD in relation to success rate and patient comfort. In addition, we show that a specialist practice dealing with a “crude” population of patients performs with a perfectly acceptable quality concerning success rate and patient comfort irrespective of use of UPD.

Furthermore, we confirm an old prejudice as we prove age and gender to be very strong discriminators for a successful colonoscopy (Table 6) [10].

CONCLUSION

A positioning device is convenient to have, but not a necessity. These devices are ”nice to have” rather than “need to have”.

However, all participating specialists except one preferred working with the UPD, and it cannot be out-ruled that there is a small and in this study insignificant tendency that the UDP makes difficult colonoscopies easier to complete.

Correspondence: Lene Hendel, Rolighedsvej 47, 3460 Birkerød, Denmark. E-mail: lene@hendel.dk

Accepted: 19 February 2013

Conflicts of interest:none. Disclosure forms provided by the authors are available with the full text of this article at www.danmedj.dk.

Acknowledgement: We are thankful for skilful expert statistical assistance from Signe Olrik Wallenstein Jensen. Furthermore, we are grateful for the assistance provided by secretaries, nurses and medical students – all from our staffs, and to Nina Kjaergaard for English proofreading.

Referencer

  1. Williams C, Guy C, Gillies D et al. Electronic three-dimensional imaging of intestinal endoscopy. Lancet 1993;341:724-5.

  2. Saunders BP, Bell GD, Williams CB et al. First clinical results with a real time, electronic imager as an aid to colonoscopy. Gut 1995;36:913-7.

  3. Syed G, Shah, Brooker JC et al. Effect of magnetic endoscope imaging on patient tolerance and sedation requirements during colonoscopy: a randomized controlled trial. Gastrointest Endosc 2002;55:832-7.

  4. Cheung HYS, Chung CC, Kwok SY et al. Improvement in colonoscopy performance with adjunctive magnetic endoscope imaging: a randomized controlled trial. Endoscopy 2006;38:214-7.

  5. Andersen FH. Colonoscopies performed as surgical outclinic procedures. Ugeskr Læger 2007;169:514-7.

  6. Waye JD, Rex DK, Williams CB. Colonoscopy: principles and practice, 2 ed. Chapter 46: magnetic imaging for colonoscopy. Wiley-Blackwell, 2009:619-26.

  7. Jess P, Bulut O, Almasi A et al. The usefulness of a magnetic endoscope locating device in colonoscopy in daily practice. Surg Endosc 2009;23:1353-5.

  8. Hoff G, Bretthauer M, Dahler S et al. Improvement in caecal intubation rate and pain reduction by using 3-dimensional magnetic imaging for unsedated colonoscopy: a randomized trial of patients referred for colonoscopy. Scand J Gastroenterol 2007;42:885-9.

  9. Dechêne A, Jochum C, Bechmann L et al. Magnetic endoscopic imaging saves abdominal compression and patient pain in routine colonoscopies. J Dig Dis 2011;12:364-70.

  10. Sint Nicolaas J, de Jonge V, Korfage IJ et al. Benchmarking patient experiences in colonoscopy using the Global Rating Scale. Endoscopy 2012;44:462-72.

  11. Harewood GC. Relationship of colonoscopy completion rates and endoscopist features. Dig Dis Sci 2005;50:47-51.

  12. Sedlack RE. Training to competency in colonoscopy: assessing and defining competency standards. Gastrointest Endosc 2011;74:355-65.

  13. Bowles CJA, Leicester R, Romaya C et al. A prospective study of colonoscopy practice in the UK today: are we adequately prepared for national colorectal cancer screening tomorrow? Gut 2004;53:277-83.

  14. Rolighed L, Andersen LM. Study of the quality of colonoscopies. Ugeskr Læger 2008;170:2232-4.